Background:
We report a case of negative serum creatinine (SCr) value due to the interference of immunoglobulin M (IgM) paraprotein.
Methods:
We measured SCr using dilution patterns of 1:10, 1:20, 1:50 on Roche Cobas®c702 analyzer. Then the patient's serum was precipitated 1:1 with polyethylene glycol (PEG) 6000 and was measured for SCr. Additionally, we used the VITROS® 350 Dry Chemistry platform to measure SCr.
Results:
The correct SCr results could not be obtained by normal and dilution patterns, and there were abnormal reaction curves. Reliable SCr results were obtained by PEG precipitation or dry chemical determination.
Conclusions:
The negative Scr value by enzymatic method was caused by the IgM paraprotein. Looking at the SCr reaction curve was an easy way to observe the presence of IgM paraprotein interference.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Serum Studies: Renal Function Tests
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...
Nephrotic Syndrome II : Assessment and Medical Management