Detection of Polyphonic Alarm Sounds From Medical Devices Using Frequency-Enhanced Deep Learning: Simulation Study
Kazumasa Kishimoto1,2,3, Tadamasa Takemura4, Osamu Sugiyama5
1Graduate School of Informatics, Kyoto University, 54 Kawara-cho, Shogoin, Sakyo-ku, Kyoto, 591-8022, Japan, 81 75-366-7701.
Background:
Although an increasing number of bedside medical devices are equipped with wireless connections for reliable notifications, many nonnetworked devices remain effective at detecting abnormal patient conditions and alerting medical staff through auditory alarms. Staff members, however, can miss these notifications, especially when in distant areas or other private rooms. In contrast, the signal-to-noise ratio of alarm systems for medical devices in the neonatal intensive care unit is 0 dB or higher. A feasible system for automatic sound identification with high accuracy is needed to prevent alarm sounds from being missed by the staff.
Objective:
The purpose of this study was to design a method for classifying multiple alarm sounds collected with a monaural microphone in a noisy environment.
Methods:
Features of 7 alarm sounds were extracted using a mel filter bank and incorporated into a classifier using convolutional and recurrent neural networks. To estimate its clinical usefulness, the classifier was evaluated with mixtures of up to 7 alarm sounds and hospital ward noise.
Results:
The proposed convolutional recurrent neural network model was evaluated using a simulation dataset of 7 alarm sounds mixed with hospital ward noise. At a signal-to-noise ratio of 0 dB, the best-performing model (convolutional neural network 3+bidirectional gate recurrent unit) achieved an event-based F1-score of 0.967, with a precision of 0.944 and a recall of 0.991. When the venous foot pump class was excluded, the classwise recall of the classifier ranged from 0.990 to 1.000.
Conclusions:
The proposed classifier was found to be highly accurate in detecting alarm sounds. Although the performance of the proposed classifier in a clinical environment can be improved, the classifier could be incorporated into an alarm sound detection system. The classifier, combined with network connectivity, could improve the notification of abnormal status detected by unconnected medical devices.
More Related Videos
04:04Asthma Detection Research Based on Voice Signal Processing and Machine Learning
Published on: July 22, 2025
06:22Machine Learning-Based Cough Tone Classification: Diagnostic Exploration of Chronic Obstructive Pulmonary Disease and Respiratory Tract Infections
Published on: September 19, 2025
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V) valves at the...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities: