Comparisons of respiratory mechanics in patients under different variations of noisy pressure support ventilation
Zhenjie Jiang1, Guixia Peng2, Yufeng Liang3
1Department of Critical Care Medicine, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou, China.
Background:
Compared with conventional pressure support ventilation (PSV), noisy PSV offers a certain degree of pressure variability, which might benefit the process of weaning from mechanical ventilation. We investigated respiratory mechanics under different levels of variation during the weaning phase in patients receiving noisy PSV to identify the optimal level of variation.
Methods:
This self-controlled before-and-after study of mechanically ventilated patients in the weaning phase was conducted from September 2020 to October 2022. Each eligible patients received noisy PSV with 0%, 15%, 25%, or 35% variation in a random order. Their respiratory mechanics were collected 1 h after the pressure variation level was switched and compared.
Results:
The study found that noisy PSV significantly improved patients' respiratory system compliance at variation levels of 25% and 35% (0% vs. 15% vs. 25% vs. 35%: 40 vs. 38.1 vs. 42.3 vs. 44.5 mL/cmH2O, P=0.005). Importantly, subgroup analysis revealed this improvement was more pronounced in chronic obstructive pulmonary disease (COPD) patients (0% vs. 15% vs. 25% vs. 35%: 40 vs. 38.4 vs. 45 vs. 50, P=0.03). At the same time, at variability levels of 25% and 35%, patients' inspiratory trigger delay time was also significantly shorter (0% vs. 15% vs. 25% vs. 35%: 140 vs. 100 vs. 100 vs. 100 ms, P<0.001). In addition, the PeakArea was significantly lower at variation 35% (8.8±2.5 cmH2O·s).
Conclusions:
Pressure variation at level of 25% or 35% could improve patients' respiratory system compliance and shorten the inspiratory trigger delay time, potentially being optimal in noisy PSV during the weaning phase from mechanical ventilation.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...