A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
One-Year Outcomes of the MODULAR ATP Trial: A Novel Leadless Pacemaker in Wireless Communication With a Subcutaneous
Michael S Lloyd1, Vivek Y Reddy2, Paul Roberts3
1Section of Cardiac Electrophysiology, Emory University, Atlanta, GA (M.S.L.).
Background:
MODULAR antitachycardia pacing (ATP), a multicenter, international trial, assesses a modular cardiac rhythm management system: a subcutaneous implantable cardioverter defibrillator in wireless communication with a leadless pacemaker (LP) capable of pace-terminating ventricular tachycardia.
Methods:
Enrolees had one or more clinical risk factors for ventricular tachycardia and did not require chronic pacing. Complications included prespecified major LP system- and procedure-related complications, and any complication related to the LP, subcutaneous implantable cardioverter defibrillator, implantation, or study protocol. Survival analysis was performed to identify complication-free rates, therapy delivery, and all-cause mortality.
Results:
The 297 patients enrolled had an ejection fraction of 35±13%, 43% secondary prevention indications, and 59% with prior ventricular arrhythmias. Of 286 patients undergoing LP implantation (100% success), 251 patients completed 12-month follow-up. Mortality rate was 6%, with none related to the implant procedure. Median follow-up duration was 23.4 months (interquartile range, 17.9-28.1). The LP major complication-free rate was 97.2%, exceeding the performance goal. The overall LP+ subcutaneous implantable cardioverter defibrillator system-related complication-free rate was 88.5%. Appropriate tachyarrhythmia-therapy (ATP+shock) rates were 14.4%, and appropriate shock rates were 8.5%. Inappropriate total tachyarrhythmia therapy was 9.5% of which 8.5% were shocks. ATP was 67.3% successful in terminating ventricular arrhythmia episodes and accelerated ventricular arrhythmias in 10.1% of episodes. Overall therapy burden (ATP+shock) was 96/100 patient-years, of which 44/100 patient-years was for shock delivery.
Conclusions:
One-year outcomes of the first modular pacing-defibrillator system reveal low system and LP complication rates and good ATP efficacy rates, suggesting that the modular cardiac rhythm management is a viable alternative to single-chamber implantable cardioverter defibrillators using low-energy pacing capability without the need for transvenous leads.
Clinical Trial Registration:
URL: https://clinicaltrials.gov/; Unique identifier: NCT04798768.