Biomechanical influence of hinge wire application and removal in medial opening wedge high tibial osteotomy: A
Takaaki Hiranaka1,2, Adam Garry Redgrift3, Adia Shadd4
1Fowler Kennedy Sports Medicine Clinic, Western University, London, Ontario, Canada.
Purpose:
To compare tensile strain changes on the lateral tibial cortex during the opening procedure of medial opening wedge high tibial osteotomy (MOWHTO) with and without hinge wire, and to evaluate strain changes after hinge wire removal.
Methods:
MOWHTO was performed on five pairs of fresh-frozen human cadaveric knees (10 knees in total), which were allocated into two groups: with a hinge wire (n = 5) and without a hinge wire (n = 5). A uniaxial strain gauge was attached to the lateral tibial cortex to measure tensile strain, with negative values indicating compression and positive values indicating tension. All procedures were guided by a 3D-printed patient-specific cutting guide with a target opening gap of 9 mm, and fixation was completed using a locking plate. Tensile strain during the opening procedure was compared between groups, and strain before and after wire removal was assessed in the hinge wire group.
Results:
In four out of five matched pairs, the hinge wire group exhibited smaller tensile strain changes during opening than the contralateral group without a hinge wire. During the opening procedure, the median tensile strain change was -267 με (range: -2390 to -213 με) in the hinge wire group and -753 με (range: -1889 to -383 με) in the without hinge wire group (p = 0.345). Following hinge wire removal, the median strain change was minimal (44 με; range: 21-72 με).
Conclusions:
The hinge wire group showed smaller tensile strain changes during the opening procedure in most specimens, though the difference was not statistically significant. These findings provide preliminary biomechanical evidence suggesting that the hinge wire may serve as a protective method against fracture of the lateral hinge during MOWHTO, and that its removal after the plate fixation does not compromise lateral cortex stability.
Level Of Evidence:
N/A.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
07:33In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Related Concept Videos
Carbon Skeletons
Bone Structure
Bone Remodeling
Gross Anatomy of Bone
The diaphysis is the tubular shaft that runs between the proximal and distal ends of the bone. The walls of the diaphysis are composed of dense and hard compact bone made of numerous osteons — the functional unit of the compact bone. The hollow region in the diaphysis is called the medullary cavity, which harbors the bone marrow. In infants and children, this marrow cavity is filled with red marrow, whereas in adults, it...
Bone Markings
Articulating Projections
Articulating projections are found where two bones meet to form a joint. These structures are usually found at the ends of bones. The largest articulation is a rounded projection called the head, supported by a narrow neck at the ends of...