Validation of a candidate international harmonisation protocol-compliant measurement procedure for reticulocyte

Yutaka Nagai1,2,3, Tomohiro Takeda2, Hiromichi Matsushita1,4

  • 1Department of Laboratory Medicine, Keio University School of Medicine, Tokyo, Japan.

PubMed

Insights

This study validates a new flow cytometry method for measuring reticulocyte percentage, confirming its accuracy and compliance with international standards. The thiazole orange/CD41a/CD61-MP offers a reliable alternative for clinical diagnostics.

Area of Science:

  • Hematology
  • Clinical Chemistry
  • Biotechnology

Background:

  • The International Harmonisation Protocol (IHP) for Reticulocyte Percentage (%Retic) relies on microscopic methods (NMB-IHP).
  • Flow cytometry (FCM) offers practical advantages for accurate reticulocyte measurement.
  • Existing FCM methods (TO/CD235a-MP) lack validation against NMB-IHP.

Purpose of the Study:

  • To validate a novel FCM-based Measurement Procedure (MP) using thiazole orange and anti-CD45/CD41a/CD61 antibodies (TO/CD41a/CD61-MP) as an IHP-compliant method.
  • To compare the TO/CD41a/CD61-MP with established NMB-IHP and TO/CD235a-MP.
  • To assess the practical utility and accuracy of TO/CD41a/CD61-MP on automated hematology analyzers.

Main Methods:

  • Validated TO/CD235a-MP against NMB-IHP.
  • Compared the novel TO/CD41a/CD61-MP with the validated TO/CD235a-MP using regression analysis.
  • Assessed the accuracy of TO/CD41a/CD61-MP on Sysmex XN-2000™ and Nihon Kohden Celltac G+™ analyzers.

Main Results:

  • TO/CD235a-MP was confirmed to be IHP-compliant.
  • Regression analysis showed strong consistency between TO/CD41a/CD61-MP and TO/CD235a-MP.
  • High correlation coefficients (0.97) were observed between TO/CD41a/CD61-MP and automated analyzers for %Retic (0.0-8.2).

Conclusions:

  • The TO/CD41a/CD61-MP is validated as an IHP-compliant method for reticulocyte measurement.
  • This study provides the first empirical validation of FCM-based MPs for reticulocyte counting.
  • The TO/CD41a/CD61-MP demonstrates practical utility and accuracy for clinical applications.