Comparing pulse oximeter performance using a common functional tester versus controlled desaturation studies on
Seif Elmankabadi1,2,3, Jake Dove4, Ella Behnke5,6
1Department of Anesthesia and Perioperative Care, Zuckerberg San Francisco General Hospital and Trauma Center, University of California San Francisco, San Francisco, CA, USA. seif.elmankabadi@ucsf.edu.
Abstract:
Functional testers are designed to evaluate select pulse oximeter characteristics but are often misused to validate device accuracy, potentially providing false reassurance. This study evaluated whether the Fluke ProSim8 (FPS8) could accurately predict oximeter performance during human controlled desaturation studies or identify performance differences under low signal conditions. 12 oximeters were tested using two FPS8 protocols: (1) an 'SpO₂ plateau' protocol which mimicked controlled desaturation studies by evaluating device performance over a range of simulated SpO2 (70-100%), and (2) a 'signal space' protocol designed to assess device accuracy under varying modulation and transmission conditions. Each device also underwent controlled desaturation testing in healthy adults. Six of the 12 oximeters passed (ARMS ≤ 3%) the SpO₂ plateau protocol; however, three of these failed (ARMS > 3%) human testing. At lower simulated saturations, most devices overestimated SpO₂. In the signal space protocol, oximeters performed well under high signal conditions, but many failed to produce readings or showed SpO₂ errors > 3% under low signal conditions. On average, oximeters failed to generate a reading 20.2 ± 7.2 times out of 60 attempts. Ten devices passed (ARMS < 3%) the signal space protocol, but two of these failed human testing. Oximeter performance on the FPS8 did not correlate with human performance (R² = 0.08 for the plateau protocol; R² = 0.01 for the signal space protocol). The FPS8 did not reliably predict oximeter accuracy in human desaturation studies or under low signal conditions; current functional tester protocols are limited in predicting real-world oximeter performance.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...