Lingual pulse oximeter probe location and weight category influence reported pulse oximeter values in healthy
Kathryn E Ryan1, Nathaniel Kapaldo1, Eduarda M Bortoluzzi2
1Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS.
Objective:
To describe contact forces of 3 pulse oximeter clips and to evaluate, in healthy anesthetized dogs and cats, if perfusion index (PI) or peripheral arterial blood oxygen saturation (SpO2) is influenced by lingual clip location or weight category.
Methods:
Benchtop investigation of 3 clips occurred in February 2024 to determine contact force with increasing opening distances. In live animals, 1 clip was evaluated at 2 lingual locations. Enrollment of client-owned dogs and cats occurred from March 13, 2024, through June 21, 2024, including 10 each of dogs < 10 kg, dogs > 20 kg, and cats. Data included SpO2, PI, and magnitude of clip-induced lingual compression.
Results:
Clips exhibited a positive relationship between opening distance and mean force generation. Clip location and weight category in dogs influenced SpO2 and PI values. At thicker tongue locations, median SpO2 values in dogs < 10 kg and cats were 5% and 2% greater, respectively, compared to thinner locations. Dogs > 20 kg displayed a median SpO2 value 1% lower at the thicker tongue location compared to the thinner location. Median PI values were lower at the thinner tongue site in dogs and cats.
Conclusions:
Pulse oximeter clip opening distance and contact force exhibited a positive relationship. Thicker lingual sites generated higher SpO2 and PI values in cats and dogs < 10 kg.
Clinical Relevance:
Pulse oximeter position at 2 different locations on the tongue influence SpO2 and PI values in healthy anesthetized dogs < 10 kg, dogs > 20 kg, and cats.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...