Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Update on first-in-human experience of high-energy ElectroPulse pulsed field ablation
Saikiran Kakarla1, Mehrdad Emami1, Alexis Puvrez1
1Centre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, South Australia, Australia.
Background:
The ElectroPulse system is a novel pulsed field ablation (PFA) platform integrating mapping and ablation in a small-diameter catheter with irrigation and wire-free design.
Objective:
The purpose of this study was to present mid-term outcomes from 2 sequential first-in-human cohorts for pulmonary vein isolation and posterior wall isolation (PVI and PWI).
Methods:
The ElectroPulse trial enrolled 60 patients with symptomatic atrial fibrillation undergoing de novo ablation. Cohort 1 (n = 30) followed the standard protocol, while cohort 2 (n = 30) used a modified protocol. The primary end point was time to first atrial arrhythmia ≥30 seconds; secondary end points included efficacy and safety, with exploratory analyses of pulmonary vein reconnection and lesion durability. Outcomes were assessed through clinic visits, electrocardiograms, and 4-day Holter monitoring every 3 months.
Results:
Compared with patients in cohort 1, those in cohort 2 were older (P = .005), had larger left atrial volumes (P = .009), and higher CHA2DS2-VA scores (P = .002). The modified protocol reduced PFA applications and procedure time (P < .001), with 100% (n = 60) acute PVI/PWI in both cohorts. At 6 months, 16% (n = 5) of cohort 1 and 13% (n = 4) of cohort 2 had atrial fibrillation recurrence (P = .834). Redo ablation was required in 16.6% (n = 5) of cohort 1 but only in 3.3% (n = 1) of cohort 2 (P = .19). Durable PVI was seen in 27% (n = 3) of repeat procedures in cohort 1 vs 100% (n = 3) in cohort 2. Complication rates were low and similar.
Conclusion:
ElectroPulse PFA enables safe and effective PVI and PWI. The updated high-energy protocol improves efficiency by reducing applications and procedure duration while increasing durability without compromising safety or efficacy.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electroconvulsive Therapy