Related Experiment Video
Updated: Jan 11, 2026

09:07
Author Spotlight: Advancing Endoscopic Ossiculoplasty – Techniques, Innovations, and Practical Guidance for Clinical Integration
Published on: January 26, 2024
3.3K
Multicenter study results of the med-el partial ossicular replacement prostheses
Susan Arndt1, Christian Offergeld1, Wendelin Wolfram2
1Department of Otorhinolaryngology-Head and Neck Surgery, Faculty of Medicine, Medical Center-University of Freiburg, University of Freiburg, 79106, Freiburg, Germany.
Summary
Partial ossicular replacement prostheses, including mXACT and mCLIP, show safe and effective outcomes. Most patients achieved successful audiological results with satisfactory short-term hearing improvement.
Area of Science:
- Otolaryngology
- Biomedical Engineering
- Audiology
Background:
- Middle ear reconstruction often involves partial ossicular replacement prostheses.
- Evaluating the safety and effectiveness of specific prostheses is crucial for clinical decision-making.
Purpose of the Study:
- To analyze the safety and effectiveness of mXACT, mXACT PRO, mCLIP, and mCLIP ARC partial ossicular replacement prostheses.
- To compare outcomes based on coupling type (bell vs. clip).
Main Methods:
- Retrospective analysis of clinical data from 203 patients undergoing partial ossicular replacement.
- Evaluation of adverse events (AEs) and audiological outcomes via pure-tone audiometry (PTA4).
- Success defined as a post-operative PTA4 air bone gap (ABG) ≤ 20 dB.
Main Results:
- Successful rehabilitation (PTA4 ABG ≤ 20 dB) was achieved in 69.8% of the bell group and 72.4% of the clip group.
- The bell group reported 5.9% AEs, while the clip group reported 11.2% AEs.
- Mean post-operative PTA4 ABG was 17.8 dB (bell) and 16.3 dB (clip), with stable bone conduction thresholds in over 95% of patients.
Conclusions:
- mXACT and mCLIP partial ossicular replacement prostheses demonstrate satisfactory short-term audiological results.
- These prostheses are considered safe and effective for partial ossicular replacement.
- The study supports the use of these devices in middle ear reconstruction.
Keywords:
MCLIP ARC partial prosthesisMCLIP partial prosthesisMXACT PRO partial prosthesisMXACT partial prosthesisPartial ossicular replacement prostheses