Smartphone-Based Blood Pressure Monitoring via the Oscillometric Finger Pressing Method: Investigation of the DC
Objective:
Oscillometric finger pressing is a potential method for smartphone-based blood pressure (BP) monitoring. A photoplethysmography (PPG)-force sensor unit measures the slowly increasing finger pressure applied by the user under visual guidance and the resulting variable blood volume oscillations ("AC PPG"). BP can then be estimated from the oscillation height versus finger pressure function. The non-oscillating component of PPG ("DC PPG") during oscillometric finger pressing was investigated.
Methods:
The total (AC+DC) PPG waveform, finger pressure, and ECG waveform during finger pressing were measured with a modified custom system along with arm cuff BP in volunteers. A mathematical model accounting for the arterial compliance curve and tissue compression was developed to explain the measured total PPG waveform versus finger pressure function. The model predicted that DC PPG (average of the total PPG over each heartbeat) versus finger pressure function will show a fiducial marker (bend) near finger systolic BP (SP). An algorithm was developed to detect this bend and estimate arm SP from the finger pressing measurements.
Results:
The model explained the measured total PPG waveform versus finger pressure function over the finger pressure range that is relevant to BP estimation. The model-based algorithm yielded a correlation coefficient of 0.90 and a precision error of 9.2 mmHg against cuff SP (N = 18).
Conclusion:
An easy-to-understand model can explain the total PPG waveform during finger pressing, and SP can be estimated from the DC (non-oscillating) PPG versus finger pressure function.
Significance:
These findings may prove useful in converting ubiquitous smartphones into BP sensors.
More Related Videos
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure