Multimodal Analytical Platform on a Multiplexed Surface Plasmon Resonance Imaging Chip for the Analysis of Extracellular Vesicle Subsets
Published on: March 17, 2023
Multimethod Platform Based on Dynamic Image Analysis and spICP-MS for Number-Based Quantification of Microplastics
Aneta Sikora1, David Ojeda1, Dorota Bartczak1
1National Measurement Laboratory, LGC Limited, The Priestley Centre, 10 Priestley Road, Guildford, Surrey GU2 7XY, United Kingdom.
A new multimethod platform combining dynamic image analysis (DIA) and single particle inductively coupled plasma mass spectrometry (spICP-MS) reliably quantifies small microplastics. This validated approach accurately determines particle number concentration for microplastic analysis.
Area of Science:
- Environmental Science
- Analytical Chemistry
- Materials Science
Background:
- Microplastic pollution is a growing environmental concern.
- Accurate quantification of small microplastics (1-10 μm) is challenging.
- Existing methods often lack the precision needed for reliable particle number concentration determination.
Purpose of the Study:
- To develop and validate a multimethod platform for reliable microplastic particle number concentration measurement.
- To assess the performance of dynamic image analysis (DIA) and single particle inductively coupled plasma mass spectrometry (spICP-MS) for microplastic quantification.
- To identify critical parameters influencing measurement accuracy and uncertainty.
Main Methods:
- Combined dynamic image analysis (DIA) and single particle inductively coupled plasma mass spectrometry (spICP-MS).
- Validated the platform using polystyrene (PS) microparticles (1-10 μm).
- Investigated instrumental parameters affecting accuracy, including detection threshold (DIA) and transport efficiency calibration (spICP-MS).
Main Results:
- The DIA/spICP-MS platform reliably detected and quantified microplastics.
- Optimal conditions yielded comparable number-concentration values for 5 μm PS-Latex microspheres.
- Measurement uncertainties were approximately 2.5% (DIA) and 10.2% (spICP-MS) under optimal conditions.
- Key uncertainty contributors identified: particle detection variability (DIA) and transport efficiency calibration (spICP-MS).
Conclusions:
- The developed DIA/spICP-MS platform offers a reliable method for determining small microplastic particle number concentration.
- Instrumental parameter optimization is crucial for accurate microplastic quantification.
- Understanding uncertainty contributions is essential for robust microplastic analysis.