"hDOS": an automated hybrid diffuse optical device for real-time noninvasive tissue monitoring: precision and in vivo
Marta Zanoletti1, Muhammad Atif Yaqub1, Lorenzo Cortese1
1ICFO-Institut de Ciències Fotòniques, The Barcelona Institute of Science and Technology, Barcelona, Spain.
Significance:
The hybrid diffuse optical system (hDOS) is a fully automated platform designed to bring advanced optical monitoring closer to clinical practice. Many existing systems lack automation, multiparametric output, and operator independence, limiting their use in demanding environments such as intensive care. hDOS integrates time-domain near-infrared spectroscopy, diffuse correlation spectroscopy, and pulse oximetry to assess both tissue oxygenation and perfusion. Although initially developed in the context of vascular occlusion tests, its modular design makes it suitable for broader applications, including trauma, surgery, anesthesia, and studies in healthy subjects.
Aim:
It aims to design and validate hDOS, focusing on precision, repeatability, and usability for peripheral microvascular monitoring in both clinical and research settings.
Approach:
Validation included test-retest measurements, a 7-month clinical deployment in the critical care, and a comparison with a commercial continuous wave NIRS device (INVOS 5100C).
Results:
The device underwent extensive validation, accumulating over 200 h of usage across measurement sessions. The system showed high precision (test-retest CV for oxygenation, for perfusion), stable long-term performance, and lower variability than INVOS. It has also detected statistically significant differences during VOTs and detected hemodynamic impairment in ICU patients ( ) compared with healthy volunteers ( ).
Conclusions:
hDOS performed well in both bench and clinical settings, offering a unique combination of parameters in a fully automated, self-contained platform.
More Related Videos
07:12Real-Time Monitoring of Neurocritical Patients with Diffuse Optical Spectroscopies
Published on: November 19, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
High-Performance Liquid Chromatography: Types of Detectors
Electronic Distance Measuring Instruments