Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Accuracy and Reduced Sleep Disruption of a Wearable Smartwatch for Nocturnal Blood Pressure Monitoring: A Validation
Yanbo Liu1, Yang Lu2, Jiabo Wu2
1Department of Healthcare, Peking Union Medical College Hospital, Beijing, China.
Abstract:
Nocturnal blood pressure is a significant predictor of cardiovascular risk, but traditional ambulatory blood pressure monitoring (ABPM) can disrupt sleep. This study aimed to validate the accuracy of the Huawei Watch D2 (WD2), a wrist-worn oscillometric device, and assess its impact on sleep quality compared to conventional ABPM. This prospective, single-center trial first validated the WD2 against a mercury sphygmomanometer in 85 participants according to ANSI/AAMI/ISO 81060-2:2018 standards in both seated and supine positions. Subsequently, the device's nocturnal accuracy and sleep impact were compared against a traditional ABPM device in 46 participants over a three-night protocol. The WD2 met all ISO accuracy standards. During nighttime monitoring, there was no significant difference between the WD2 and ABPM for mean systolic or diastolic blood pressure. However, the number of awakenings was significantly lower on nights with the WD2 alone compared to nights with the ABPM device (p = 0.016). In conclusion, the Huawei Watch D2 is a clinically validated device that provides nocturnal blood pressure readings comparable to traditional ABPM with the significant advantage of minimal sleep disruption, positioning it as a valuable alternative for nocturnal hypertension monitoring.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement