The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Late-Onset Pericardial Effusion and Device Extrusion After Left Atrial Appendage Occluder Implantation
Omar Darwish1, Sai Nikhila Ghanta2, Sandeep Randhawa2
1Department of Internal Medicine, University of Arkansas for Medical Sciences, Little Rock, Arkansas, USA.
Background:
We present a case of Watchman device extrusion from left atrial appendage (LAA) free wall 6 months after implantation in a patient with longstanding persistent atrial fibrillation.
Case Summary:
A 74-year-old male with longstanding persistent atrial fibrillation of 15 years duration was implanted with a Watchman device as he cannot be on long-term oral anticoagulation due to a history of multiple falls. Six months post implantation, repeat transesophageal echocardiogram revealed a thrombus on the epicardial aspect of LAA and a large posterior pericardial effusion. The patient subsequently underwent surgical removal of both the LAA and the protruding Watchman device.
Discussion:
In the case we present, the patient developed a clinically significant pericardial effusion 6 months after the Watchman device implantation, accompanied by left atrial extrusion, which was not reported to our knowledge.
Take-Home Message:
Complications of Watchman devices respect no timeline boundaries and should be kept in mind when evaluating cardiac tamponade in patients with an implanted Watchman device.