Comparison of airway opening pressure measurements: low-constant-flow vs. constant low-slope pressure ramp methods
Ken Tonai1, Atsuko Shono2, Ryuichi Nakayama2
1Division of Intensive Care, Department of Anesthesiology and Intensive Care Medicine, Jichi Medical University School of Medicine, 3311-1 Yakushiji, Shimotsuke, Tochigi, 329-0498, Japan. k-tonai@jichi.ac.jp.
Background:
The airway closure phenomenon occurs when the airway collapses, isolating the proximal airway from the distal alveoli. Airway opening pressure (AOP) is required to reopen closed airways. Two methods are available to measure AOP: the low-constant flow (AOPflow) and constant low-slope pressure ramp methods (AOPpres). The discrepancies between these two methods remain unclear. We investigated whether there is a difference between AOPflow and AOPpres when used for mechanically ventilated patients.
Methods:
In this single-center retrospective observational study, we included 42 patients who were mechanically ventilated owing to respiratory failure in the intensive care unit of a university hospital between January 2023 and October 2024. AOP was measured using two methods: AOPflow (5 L/min) and AOPpres (2 cmH2O/s). Agreement and correlation between the two methods were evaluated using Bland-Altman plots, Passing-Bablok regression, and Spearman's rank correlation.
Results:
AOP measured using AOPflow (median: 4.7 cmH2O) was higher than that measured using AOPpres (median: 1.9 cmH2O, P < 0.001). Nevertheless, the two were strongly correlated (ρ = 0.86, P < 0.001) in all patients. The regression equation was y = 1.39x + 0.90 (95% confidence interval [CI] for slope b: 1.20-1.65, 95% CI for intercept a: 0.48-1.64). In patients with AOPflow ≥5cmH2O, AOPflow was moderately correlated with AOPpres (ρ = 0.77, P < 0.001). The regression equation was y = 0.82x + 4.51 (95% CI for slope b: 0.50-1.17, 95% CI for intercept a: 2.57-6.72). The rate of pressure increase from the pressure at the beginning (0 cmH2O) of inflation up to AOPflow was 26.4 cmH2O/s (10.5-30.0) in all patients and was moderately correlated with the difference between AOPflow and AOPpres (ρ = 0.61, P < 0.001).
Conclusions:
Systematic biases were observed between AOPflow and AOPpres, with AOPflow tending to yield higher values. However, the physiological significance of the AOP values obtained from each method remains unclear, and caution is needed for clinical application.
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...