Related Experiment Video
Updated: Jan 10, 2026

23:33
The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
84.3K
Comparing Effectiveness and Safety of Left Atrial Appendage Closure Devices: A Network Meta-Analysis of Randomized
John W Davis1, Steven L Mai2, Wissam Harmouch2
1Department of Internal Medicine, Washington University in St. Louis/Barnes-Jewish Hospital, St. Louis, Missouri, USA.
Clinical Cardiology
|November 22, 2025
Summary
The Amplatzer Amulet and Watchman left atrial appendage closure (LAAC) devices show similar risks for stroke, death, and device embolization. However, Amulet has a higher risk of pericardial effusion compared to Watchman.
Area of Science:
- Cardiology
- Medical Devices
- Interventional Cardiology
Background:
- Atrial fibrillation-related stroke is a significant cause of death and disability.
- The comparative effectiveness and safety of different left atrial appendage closure (LAAC) devices, and their comparison to anticoagulation, require further clarification.
Purpose of the Study:
- To conduct a systematic review and network meta-analysis (NMA) comparing the Watchman and Amplatzer Amulet LAAC devices.
- To compare LAAC devices against each other and against warfarin or direct oral anticoagulants (DOACs).
- To evaluate primary effectiveness outcomes (stroke, all-cause death) and safety outcomes (thromboembolism, device embolization, pericardial effusion).
Main Methods:
- Systematic review and network meta-analysis (NMA) of randomized controlled trials (RCTs).
- Included 6 RCTs with 3666 participants comparing Watchman and Amulet LAAC devices, warfarin, or DOACs.
- Primary comparisons focused on LAAC devices, with secondary comparisons to anticoagulation.
Main Results:
- No significant difference in stroke risk between Amulet and Watchman (RR=1.48, 95% CI: 0.64-3.46).
- No significant difference in all-cause death risk between Amulet and Watchman (RR=1.00, 95% CI: 0.59-1.70).
- Higher risk of pericardial effusion with Amulet compared to Watchman (RR=2.12, 95% CI: 1.45-3.09).
Conclusions:
- The Amplatzer Amulet and Watchman LAAC devices demonstrate comparable risks for stroke, mortality, thromboembolism, and device embolization.
- The risk of pericardial effusion is significantly higher with the Amulet device than with the Watchman device.
- Both LAAC devices showed an increased risk of pericardial effusion compared to warfarin.