Long-term low-voltage impedance measurements in subcutaneous implantable cardioverter-defibrillators
Giacomo Mugnai1, Luca Tomasi1, Luca Ottaviano2
1Electrophysiology and Cardiac Pacing, Division of Cardiology, Department of Cardio-Thoracic, School of Medicine, University Hospital of Verona, Verona, Italy.
Background:
High-voltage impedance (HVI), measured during subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation via defibrillation testing or a synchronized shock, is associated with defibrillation efficacy. Recently, S-ICD systems have been upgraded to automatically perform long-term measurements of low-voltage impedance (LVI) using a 1-V subthreshold pulse.
Objective:
This study evaluated LVI as a surrogate for HVI and described its long-term trends in S-ICD recipients.
Methods:
We analyzed data from 1226 patients who underwent de novo S-ICD implantation across 15 Italian centers. Weekly averages of LVI and HVI were calculated, and agreement between simultaneous measurements was assessed.
Results:
Over a median follow-up of 37 months (interquartile range 20-57 months), 373 paired HVI and LVI measurements were analyzed. LVI strongly correlated with HVI (r = 0.90; 95% confidence interval 0.88-0.92; P < .001), with a mean bias of -3 Ω (limits of agreement -21 to 14 Ω). The mean LVI increased significantly during the first 3 months postimplantation (from 59 ± 14 to 76 ± 16 Ω; P < .001) before stabilizing (77 ± 17 Ω; P = .231). Higher LVI values were observed in overweight/obese patients when subcutaneous device positioning and the 3-incision lead deployment technique were used. Similarly, higher values were obtained when significant subcoil fat was observed and the Prospective, RAndomizEd comparison of subcuTaneOus and tRansvenous ImplANtable cardioverter-defibrillator therapy score was ≥90.
Conclusion:
LVI showed strong agreement with HVI during follow-up, supporting its potential use as a noninvasive surrogate for HVI. LVI increased during the initial postimplantation period and subsequently stabilized, possibly reflecting physiological changes.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...