Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Respiratory function monitors (RFMs) used for newborn resuscitation: accuracy and performance in the presence of
Stephanie Morakeas1,2, Viktoria Gruber3,4, Murray Hinder5
1Neonatal Intensive Care Unit, Westmead Hospital, Sydney, Australia.
Background:
Use of respiratory function monitors (RFMs) during newborn resuscitation is not currently recommended by ILCOR. New-generation compact monitors have recently entered the market with portable designs offering better integration into clinical practice. Consensus on science regarding RFM use has been gained using complex legacy or research-only devices. This study aims to compare the accuracy and performance under leak conditions of two legacy RFMs with two new-generation monitors.
Methods:
The accuracy of new-generation RFMs (ResusRight-Juno, Monitvent-Neo100) were compared to legacy RFMs (Acutronics-Florian, Phillips-NM3) in the presence of imposed dynamic leaks (0 %, 20 %, 50 %, 90 %). Manual inflations using a Self-inflating bag (SIB) and T-piece Resuscitator (TPR) were delivered to a test lung (compliance 0.6 and 2 mL/cmH2O). Tidal volume, leak and pressure were measured proximal to leak by each RFM and related to traceable reference values.
Results:
Analysis of 1,920 inflations was performed. At zero leak, all RFMs performed with comparable accuracy for tidal volume, leak and pressure with minimal differences to reference measurements. System leak was accurately determined with tested RFMs at all levels. At 90 % system leak, RFM measured SIB-delivered volumes decreased significantly to <50 % compared with no leak; TPR-delivered volumes were less impacted and more accurate.
Conclusion:
The new-generation RFMs demonstrated comparable accuracy and performance to legacy devices at zero leak. All RFMs displayed accurate leak levels including 90 %. RFMs used with SIBs showed significantly reduced performance and accuracy at 90 % leak in contrast to TPR at 90% leak. New-generation RFMs simplified user interfaces may enhance benefits during clinical use.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...