Lung function assessment by electrical impedance tomography among obese patients
Wang Chun Kwok1,2, Adrien Touboul3, Wing Sun Chow4
1Department of Medicine, School of Clinical Medicine, Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong, China. kwokwch@hku.hk.
Electrical Impedance Tomography (EIT) shows promising results for measuring lung function in obese individuals. This study found EIT comparable to spirometry, suggesting its potential as a valuable tool for assessing respiratory health in obesity.
Area of Science:
- Pulmonary Medicine
- Medical Imaging Technology
- Obesity Research
Background:
- Obesity is a global health concern linked to compromised respiratory function due to mechanical factors and comorbidities like asthma.
- Spirometry, the standard for lung function assessment, has limitations including measuring only global lung function and requiring significant patient effort.
Purpose of the Study:
- To compare and correlate lung function parameters measured by Electrical Impedance Tomography (EIT) and spirometry in obese and non-obese Chinese subjects.
- To evaluate the potential of EIT as a tool for lung function measurement in obese populations.
Main Methods:
- A prospective cross-sectional study was conducted with 42 obese and 35 non-obese Chinese subjects.
- Obesity was defined using the Asian population BMI cut-off of 25 kg/m² or higher.
- Lung function parameters were measured using both spirometry and EIT, with data analyzed for correlation.
Main Results:
- Spirometric values for Forced Expiratory Volume in 1 second (FEV1) and Forced Vital Capacity (FVC) showed good correlation between spirometry and EIT in both obese and non-obese groups.
- Pearson correlation coefficients for FEV1 were 0.705 (obese) and 0.673 (non-obese), both with p < 0.01.
- Pearson correlation coefficients for FVC were 0.568 (obese) and 0.541 (non-obese), both with p < 0.01.
Conclusions:
- EIT technology demonstrates comparable performance to spirometry in measuring lung function across obese and non-obese individuals.
- EIT presents a potential alternative tool for lung function assessment, particularly beneficial for obese subjects who may find spirometry challenging.