Related Experiment Video
Updated: Jul 3, 2026

Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
High-Throughput Quantitation of Plasma Trimethylamine N-oxide Using Desorption Electrospray Ionization Mass
Kai-Yuan Chiu1,2, Yun-Chen Hsieh1, Hsin-Bai Zou3
1Department of Chemistry, National Taiwan University, Taipei 10617, Taiwan.
Abstract:
Trimethylamine N-oxide (TMAO) is an emerging biomarker of cardiovascular disease (CVD) risk, but current detection methods are limited by low throughput and lengthy workflows. To address this, we developed a high-throughput desorption electrospray ionization-mass spectrometry (DESI-MS) platform for rapid and accurate quantitation of TMAO in plasma. The method involves protein removal, spot deposition, and DESI-MS analysis using isotope-labeled internal standards for calibration. Validation showed strong linearity (R2 > 0.97), precision (CV < 20%), minimal matrix effects, and low carry-over (<5%). In a cohort of 197 patients from National Taiwan University Hospital, DESI-MS demonstrated high correlation with LC-MS/MS (R = 0.96), 92.9% concordance in risk classification, and a 10-fold reduction in processing time. Risk stratification revealed a 1.55-fold higher prevalence of coronary stenosis in the high-risk group. Capable of processing up to 2,000 samples per day, this DESI-MS platform shows strong potential for large-scale clinical screening and personalized cardiovascular risk assessment.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
04:19Author Spotlight: An Improved Technique for Trimethylamine Detection in Animal-Derived Medicine by Headspace Gas Chromatography-Tandem Quadrupole Mass Spectrometry
Published on: March 10, 2023
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
High-Performance Liquid Chromatography: Types of Detectors
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Acute Coronary Syndrome III: Diagnostic Studies