Deep learning-based reconstruction improves image quality in low-dose head CT angiography
Xin Huang1, Jin Shang1, Yao Xiao1
1Department of Medical Imaging, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, Shaanxi, 710061, China.
Objective:
This study aimed to compare the image quality of filtered back projection (FBP), adaptive statistical iterative reconstruction-Veo (ASIR-V) and the deep learning image reconstruction (DLIR) algorithms in low-dose head CT angiography (CTA).
Methods:
This prospective study was conducted on 25 patients undergoing head CTA using a 256-slice CT scanner. Patients received 25 mL of iodine contrast (Iopromide, 370 mg I/mL, 3.0 mL/s). Images were reconstructed using DLIR with high settings (DLIR-H) and medium settings (DLIR-M), FBP, and ASIR-V with a blending factor of 50% (ASIR-V 50%). CT values, standard deviations, signal-to-noise ratios (SNR), and contrast-to-noise ratios (CNR) were measured at the basal ganglia, posterior cranial fossa, center of semiovale, and middle cerebral artery. The edge rise slope (ERS) of the middle cerebral artery rim was measured to assess vessel clarity. Image noise, vessel edge definition, and overall quality were scored on a 5-point scale, while sharpness and clarity were rated on a 4-point scale.
Results:
FBP images exhibited the highest image noise, as reflected by SD values. DLIR, especially DLIR-H, showed superior noise reduction compared to ASIR-V 50%. SNR followed this trend: FBP < ASIR-V 50% < DLIR-M < DLIR-H. Spatial resolution, measured by ERS for vessel wall clarity, was higher in DLIR images compared to in ASIR-V 50%. DLIR outperformed conventional iterative algorithms in balancing noise reduction and edge clarity, with both DLIR-M and DLIR-H achieving better subjective scores for noise, edge definition, and sharpness than ASIR-V 50% and FBP.
Conclusion:
DLIR in low-dose head CTA could reduces image noise, preserve natural texture, and enhance image clarity compared with ASIR-V and FBP methods.
More Related Videos
14:08Automated Midline Shift and Intracranial Pressure Estimation based on Brain CT Images
Published on: April 13, 2013
13:16Computed Tomography and Optical Imaging of Osteogenesis-angiogenesis Coupling to Assess Integration of Cranial Bone Autografts and Allografts
Published on: December 22, 2015
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography