Automated measurement of spinopelvic alignment parameters using a spine planning software: a validation study
Ahmad Almahozi1, Anton Früh1, Tarik Alp Sargut1
1Department of Neurosurgery, Charité - Universitätsmedizin Berlin, Corporate member of Freie Universität Berlin, Humboldt-Universität Zu Berlin, and Berlin Institute of Health, Berlin, Germany.
Spine Deformity
|November 27, 2025
Summary
Brainlab Elements Spine Planning software offers fast and reliable automated spinopelvic alignment measurements for adult spinal deformities. However, it frequently fails to automatically measure key parameters like sagittal vertical axis (SVA), requiring further software development.
Area of Science:
- Spine surgery
- Medical imaging analysis
- Orthopedic biomechanics
Background:
- Accurate spinopelvic alignment assessment is critical for managing adult spinal deformities.
- Current manual measurement methods can be time-consuming and prone to variability.
- Validation of automated software tools is essential for improving clinical workflows.
Purpose of the Study:
- To validate the Brainlab Elements Spine Planning software for automated spinopelvic parameter measurement.
- To compare the accuracy and reliability of automated measurements against manual methods.
- To assess the time efficiency of the automated software.
Main Methods:
- Spinopelvic parameters were measured manually and using the Brainlab software in 21 patients with degenerative spinal disease.
- Accuracy, intraobserver, and interobserver reliability were assessed using Bland-Altman plots and intraclass correlation coefficients (ICCs).
- Measurement times for both methods were compared.
Main Results:
- The software demonstrated high reliability (ICC=1), with strong agreement with manual measurements.
- Automated measurements were significantly faster (62s vs. 227s, p<0.001).
- However, the software failed to automatically measure at least one parameter in 76.2% of cases, particularly the sagittal vertical axis (SVA) and coronal parameters.
Conclusions:
- The Brainlab Elements software provides accurate, reproducible, and time-efficient measurements for parameters it successfully captures.
- The software's current limitations in automatically measuring SVA and coronal parameters necessitate further refinement.
- Despite limitations, the software shows promise for streamlining spinopelvic alignment assessment in clinical practice.
Keywords:
Automated measurementBrainlab ElementsSpinal deformitySpine planning softwareSpinopelvic alignmentValidation study