Even-Number Bias in Pulse Oximetry Measurements in Hospitals in Denmark and the United Kingdom: An Observational
Bartholomew Ronan O'Driscoll1, Nawar Diar Bakerly2,3, Markus Harboe Olsen4,5
1Department of Respiratory Medicine, Salford Care Organisation, Northern Care Alliance NHS Foundation Trust, Salford, UK, ronan.o.driscoll@talktalk.net.
Introduction:
It is not known if pulse oximetry oxygen saturation measurements (SpO2) are subject to even-number bias and boundary effects which have been reported for other physiological measurements.
Methods:
We reviewed three large SpO2 datasets from two countries for even-number bias: 3,351,981 continuous automated SpO2 entries in Denmark (DK-continuous), 7,602,352 manual SpO2 entries in Denmark (DK-manual), and 949,718 manual SpO2 entries in the UK (UK-manual).
Results:
Even-numbered SpO2 entries were commoner than odd numbers in all three datasets. Even numbers accounted for 50.6% of DK-continuous entries (95% CI: 50.6-50.7), 57.5% of DK-manual entries (95% CI: 57.4-57.6), and 59.7% of UK-manual entries (95% CI: 59.6-59.8): DK-manual vs. DK-continuous, relative risk [RR]: 1.14, 95% CI: 1.13-1.14, p < 0.0001; UK-manual vs. DK-continuous RR 1.18, 95% CI: 1.18-1.18, p < 0.0001. The even-number bias was more pronounced in UK medical patients for whom 64.9% of SpO2 entries had even numbers (UK-manual medical entries vs. DK-continuous medical entries RR 1.29, 95% CI: 1.29-1.30, p < 0.0001). The proportion of even SpO2 numbers was particularly high (67.0%; 95% CI: 66.5-67.5%) for UK medical patients who were using supplemental oxygen.
Conclusions:
We identified even-number bias in routine pulse oximetry entries in the UK and Denmark. The bias was substantially greater for manually collected data than for automated continuous data. The high degree of even-number bias in patients who were receiving supplemental oxygen suggests that the SpO2 might be rounded to meet target saturation range or Early Warning Score boundaries. This bias could influence patient management and clinical outcomes.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring pulse
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Bias in Epidemiological Studies