Dosimetric evaluation of magnetic resonance imaging based synthetic computed tomography for head and neck photon and
Marte Kåstad Høiskar1, René M Winter1, Natalie Hornik1,2
1Department of Physics, Norwegian University of Science and Technology, Trondheim, Norway.
Background:
Accurate target delineation is essential for precise delivery of proton therapy. A magnetic resonance imaging (MRI)-only radiotherapy workflow may improve delineations, and thus improve proton therapy, but requires reliable synthetic computed tomography (sCT) for accurate dose calculations.
Purpose:
The aim was to evaluate the dosimetric accuracy of commercial software-based sCT for proton dose calculation for head and neck cancer (HNC), and benchmark it against the accuracy of photon dose calculation on the same sCT.
Methods:
MRI and planning CT (pCT) were acquired for 20 HNC patients before receiving photon therapy. A commercial software created MRI-based sCTs, which were co-registered to pCTs. For each patient, photon and proton plans based on pCT, prescribing 68 Gy to the target, were created in RayStation and recalculated on sCT. Dose-volume histogram (DVH) metrics and gamma index (2%/2 mm criteria) were analyzed, comparing pCT- and sCT-calculated dose. Local and global gamma index were calculated for low and high-dose thresholds. A gamma pass rate (GPR) was calculated for each plan and gamma index type.
Results:
For photon and proton plans, the median difference in DVH metrics between pCT- and sCT-calculated dose was < 0.4 Gy with an interquartile range < ± 0.7 Gy for all structures and metrics, except for mean dose to oesophagus (median = -0.3 Gy, range: -2.0 Gy to 0.7 Gy), oral cavity (median = 1.8 Gy, range: 0.6 Gy to 3.0 Gy), and larynx (median = -0.1 Gy, range: -1.5 Gy to 0.6 Gy) for proton plans. The median GPR for photon plans was > 97.7%, while for proton plans it was > 93.3% except for body GPR with low-dose threshold.
Conclusion:
While sCT seems feasible for photon therapy, poorer agreement between pCT- and sCT-calculated proton dose was found. Patient positioning differences between CT and MRI may partly explain the discrepancies between pCT- and sCT-calculated proton dose.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...