Adaptation and validation of a collaborative robotic equipment applied to pediatric neurosurgery
Allison Roberto da Silva1, Hélio Rubens Machado2, Marcelo Volpon Santos3
1University Hospital, Ribeirão Preto Medical School, University of São Paulo, São Paulo, Brazil. alrsilva@hcrp.usp.br.
Introduction:
The first robotic surgery in history took place in 1987 using the PUMA 200 robotic arm. Since then, several devices have been developed, but widespread availability is still hindered by their high cost, especially in public funded hospitals.
Objective:
Conception and validation of a collaborative robotic equipment for the neurosurgical treatment of pharmacoresistant epilepsy.
Methods:
This is a methodological study in which an industrial equipment was adapted and validated for neurosurgical application. Realistic tests and simulations were performed on complex, patient-specific, three-dimensional models, both in the laboratory and in the surgical environment. Such validation followed the criteria established by the FDA manual for certification of medical devices.
Results:
Precision tests analyzed the accuracy between the programmed and executed data, resulting in a mean error at the entry point of 2.21 mm and a target error of 4.41 mm.
Discussion:
Comparative studies have pointed out the advantages of robotic over conventional surgeries, such as lower risk of bleeding, reduced surgical time, faster recovery, and excellent precision. Nevertheless, commercial challenges remain, mostly related to the high costs of software, hardware, and maintenance of currently used robots. The device described herein could be an affordable option.
Conclusion:
The KUKA IIWA SERIES® robot was adapted after successful simulations in the laboratory and operation room. The equipment was successfully tested, with trajectory errors within an acceptable margin, thus becoming a viable addition to enhance surgical precision. Future steps include approval by the respective government agency and further application in real operations.