Point of Care Transcranial Color-Coded Duplex Ultrasound of the Middle Cerebral Artery
Published on: August 9, 2024
Measurement of Systolic Blood Pressure Using POCUS With Color Doppler Compared to with an Intraarterial Line
Henry Mayo-Malasky1, Daniel Ying1, Alekhya Bukkuri2
1Division of Pulmonary, Critical Care, and Sleep Medicine, LIJ/NSUH Northwell, Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Hempstead, NY USA.
Background:
In some clinical circumstances, it may be difficult to accurately measure systolic blood pressure (SBP) using direct auscultation technique or an automated oscillometric cuff pressure device. As an alternative method, this study compared the measurement of SBP using point of care ultrasound (POCUS) with color Doppler to the measurement of SBP using an intraarterial catheter.
Methods:
Study subjects were 50 patients in an intensive care unit who had an intraarterial catheter placed for monitoring blood pressure. The intraarterial catheter systolic pressure was recorded and compared to the contemporaneous measurement of SBP using POCUS with color power Doppler (CPD). The operator placed the Doppler sample volume over the brachial artery with ipsilateral inflation of a prepositioned upper arm blood pressure cuff that was inflated sufficiently to ablate blood flow in the target artery. The blood pressure cuff was then deflated until there was return of CPD signal in the brachial artery. At this moment, the corresponding blood pressure was noted on a sphygmomanometer attached to the blood pressure cuff. The values of the two methods were compared using standard statistical technique.
Results:
The intraarterial systolic pressures and CPD systolic pressures by POCUS were well correlated with a Pearsons correlation coefficient of 0.96. Bland-Altman analysis of bias and limits of agreement indicated that the POCUS with CPD measurement was sufficiently accurate to have clinical utility.
Conclusions:
The use of POCUS with CPD to measure SBP may have utility in situations where direct auscultation or automated oscillometeric cuff pressure measurements may be unreliable.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.