Phantom evaluation of involuntary mis-centering in CT Scan : consequences for radiation dose and image quality
Hassan Ou Hadda1, Mustapha Zerfaoui1, Karim Bahhous2
1LPTPME, Faculty of Sciences, University Mohammed First, Oujda, Morocco.
Purpose:
This study investigates how involuntary patient mis-centering affects dose distribution and image quality in computed tomography (CT), with the goal of reducing radiation exposure while preserving diagnostic performance.
Methods:
Mis-centering was performed by shifting a Sun Nuclear CTDI phantom vertically and laterally, with dose recorded in specific phantom holes using a 10X6-3CT ionization chamber (Radcal). Measurements were performed on three CT scanners (Philips, FUJIFILM, Hitachi) under identical acquisition parameters. Additionally, a Philips system paired with a Catphan-503 phantom was used to assess image-quality changes. Dose ratios, calculated from absorbed dose measurements in multiple phantom holes, quantified the effect of off-center positioning.
Results:
Peripheral doses were highly sensitive to displacement: a vertical offset above the isocenter reduced the dose by up to 35% at the point above the isocenter, while an increase was observed at symmetrical points, while a lateral offset reduced it by up to 18% at points in the direction of displacement. Image-quality metrics were affected to a lesser degree, likely because modern reconstruction algorithms partially compensate for mis-centering.
Conclusion:
These findings suggest that deliberate mis-centering may be considered during follow-up CT examinations to spare radiation-sensitive regions without clinically significant loss of image quality.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Computed Tomography