Step-by-step assessment of MRI-only workflow in pelvic radiotherapy: feasibility and practical implementation
1Department of Radiation Oncology, Presicion Radiation in Oncology Key Laboratory of Sichuan Province, Sichuan Cancer Hospital & Institute, University of Electronic Science and Technology of China, Chengdu, China.
Purpose:
Magnetic resonance imaging-only (MRI-only) radiotherapy workflow capitalizes on the superior soft tissue contrast of MRI while eliminating computed tomography-MRI (CT-MRI) registration uncertainties. Most existing studies focus on optimizing individual components of this workflow, i.e., synthetic CT (sCT) generation, dose calculation accuracy, and the reliability of MRI-based patient positioning, while often neglecting comprehensive evaluation of the whole clinical process and its interdependent technical requirements.
Methods:
A total of 55 pelvic cancer patients, who underwent the standard radiotherapy workflow based on CT-MRI co-registration and involving a variety of imaging scanners, treatment planning systems (TPSs), and linear accelerators (LINACs), were included. For each patient, a fully integrated MRI-only approach was developed through a step-by-step evaluation of the different radiotherapy phases, including MRI-alone delineation, sCT generation with dose calculation, and cone beam CT-MRI (CBCT-MRI)-guided positioning. Comparative evaluations with the standard workflow were performed for target segmentation (dice similarity coefficient [DSC] and Hausdorff distance [HD]), dose calculation (DVH parameters and gamma analysis), and daily patient positioning (CBCT-CT versus CBCT-MRI registrations).
Results:
The proposed MRI-only workflow is applicable throughout all radiotherapy phases and compatible across multiple imaging systems, treatment planning systems, and linear accelerators. MRI-only delineation showed excellent agreement for the prostate (DSC = 0.948 ± 0.028) and cervix (DSC = 0.940 ± 0.015), whereas contour agreement for the rectum and prostate bed (DSC = 0.696 ± 0.037 and 0.689 ± 0.047, respectively) fell within the range of inter-observer variability. Dosimetric comparisons revealed no significant differences between CT- and sCT-based plans (p > 0.05), with median gamma pass rates of 99.0% and 97.9% for the whole body using 3%/3 mm and 2%/2 mm criteria, respectively. CBCT-MRI registration indicated positioning errors comparable to CBCT-CT.
Conclusion:
MRI-only pelvic radiotherapy workflows are clinically feasible through customized protocols for target and OAR segmentation, MRI scanner setup and immobilization, sequence selection, sCT-based dose calculation validation, and CBCT-MRI matching reliability.
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...