Comparing Lower- and Higher-Order Aberrations: Zywave® II Hartmann-Shack Wavefront Aberrometer versus Peramis
Siamak Zarei-Ghanavati1,2, Mojtaba Abrishami1,3, Maryam Hedayati1
1Eye Research Center, Department of Ophthalmology, Khatam-Al-Anbia Eye Center, Mashhad University of Medical Sciences, Mashhad, Iran.
Purpose:
To assess the agreement between Zywave® II and Peramis CSO aberrometers for lower- and higher-order aberration (LOA and HOA) measurements on dynamic conditions without cycloplegia.
Methods:
In this prospective comparative study, participants aged 20-45 years were examined. Exclusion criteria included previous ocular surgery or trauma, recent contact lens wear, and any ocular or systemic diseases. Each device was operated by an experienced operator who remained blind to the data obtained from the other aberrometer. We compared LOA measurements and the root mean square (RMS) of coma, spherical aberration, and total third- and fourth-order HOAs between the two devices, and the optical zone for measuring HOAs was the same in both the devices.
Results:
In the study involving 42 eyes of 21 participants (52.4%, female), excellent agreement was observed in LOAs (sphere and cylinder) for both the right and left eyes, with intraclass correlation coefficients of 0.96 and 0.95, respectively, using a 6 mm pupil. In addition, good-to-excellent reliability was reported for the agreement between the two devices in total HOA (t.HOA) and the RMS error of the total aberration, for pupil sizes of 5 mm and 6 mm. However, there was poor agreement between the two devices for third- and fourth-order aberrations in both the pupil sizes.
Conclusions:
The present study demonstrates agreement between the Zywave® II and Peramis® aberrometers in measuring sphere, cylinder, t.HOA, and total aberration. Nevertheless, notable differences were identified in third- and fourth-order aberrations, suggesting that specific measurements may not consistently align between devices, and these values should not be considered interchangeable.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Interval Level of Measurement
Data measured using the interval scale are similar to ordinal level data because they have a definite arrangement. However, in the interval level of measurement, the differences between data values are meaningful even though the data does not have a starting point.
Temperature is measured using the interval scale. It is measurable data, and the difference between...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Wald-Wolfowitz Runs Test II
For binary data, runs are identified using symbols such as + and −, or equivalently, 1s and...
Electronic Distance Measuring Instruments
Differential Leveling