Device-Specific Calibration Methods for Optical Frequency Domain Reflectometry-Based Shape Sensing in Catheters and
Summary
Custom calibration of fiber optic sensors improves accuracy in minimally invasive procedures. This enhances precision for vascular catheters and prostate surgical needles, reducing reliance on traditional imaging and improving patient outcomes.
Area of Science:
- Medical Devices
- Biomedical Engineering
- Sensing Technologies
Background:
- Minimally invasive procedures for arterial diseases and prostate cancer face navigation and precision challenges.
- Traditional imaging methods have accuracy limitations in complex anatomies.
- Fiber optic sensors with optical frequency domain reflectometry (OFDR) offer potential solutions.
Purpose of the Study:
- To develop and assess custom calibration techniques for fiber optic sensors in vascular catheters and prostate surgical needles.
- To address device-specific characteristics affecting shape sensing accuracy.
- To evaluate the impact of calibration, tool characteristics, and spatial resolution on shape reconstruction.
Main Methods:
- Developed custom calibration protocols for fiber optic sensors integrated into vascular catheters and prostate surgical needles.
- Assessed shape reconstruction accuracy based on calibration methods and tool characteristics.
- Quantified root-mean-squared-error (RMSE) for both catheter and needle calibration.
Main Results:
- Catheter calibration yielded an RMSE of 1.67 ± 0.77 mm (0.4% ± 0.2%).
- Needle calibration achieved an RMSE of 0.23 ± 0.14 mm (0.2% ± 0.1%).
- Spatial resolution impact was not significant but noted as application-dependent.
Conclusions:
- Custom calibration significantly enhances the accuracy and reliability of fiber optic shape sensing for minimally invasive tools.
- These methods improve safety and precision, reducing dependence on imaging like fluoroscopy.
- Potential for automation exists, promising improved clinical outcomes and workflow efficiency.