Performance of optically tracked instrumented pointers: a systematic literature review
Kristóf Rácz1, Beáta Seregély2, Rita M Kiss3
1Department of Mechatronics, Optics and Mechanical Engineering Informatics, Faculty of Mechanical Engineering, Budapest University of Technology and Economics, Műegyetem rkp. 3., H-1111, Budapest, Hungary. racz.kristof@mogi.bme.hu.
Optically tracked pointers in motion analysis can have errors up to 5 mm, while surgical navigation pointers aim for sub-millimeter accuracy. Key performance factors include tracking volume, instrument design, and tip calibration for improved localization systems.
Area of Science:
- Biomedical Engineering
- Medical Physics
- Kinesiology
Background:
- Systematic review on optically tracked instrumented pointers for motion analysis.
- Includes surgical navigation research for comparative insights.
Purpose of the Study:
- Overview of pointer performance in motion analysis.
- Establish benchmarks for localization systems.
Main Methods:
- Searched Scopus, ScienceDirect, PubMed, Web of Science up to April 2025.
- Keywords: 'pointer', 'anatomical landmark', 'motion analysis'.
- Included 24 papers (16 surgical navigation, 8 motion analysis).
Main Results:
- Motion analysis pointers: errors up to 5 mm.
- Surgical navigation pointers: expected errors < 1 mm.
- Critical factors: tracking volume, instrument length, marker arrangement, tip calibration.
Conclusions:
- Performance benchmarks for current and future localization systems.
- Informs users for system comparison and developers for accuracy improvements.
- Highlights importance of calibration and design in pointer accuracy.
More Related Videos
05:26Author Spotlight: A Streamlined and Accessible Analysis Method to Quantify Optokinetic Reflex Tracking Responses
Published on: April 12, 2024
10:25Technical Approach for Infrared Tracking for Soft Tissue Navigation with a Holographic Head-Mounted Display and Preclinical Validation
Published on: September 2, 2025
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Electronic Distance Measuring Instruments