Related Experiment Video
Updated: Jan 6, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Early detection of postoperative infections using continuous temperature monitoring: A prospective clinical trial
Lars Schäfer1, Franziska Dickel2, Karl Strohmayer3
1Department of Surgery, Medical University of Graz, Graz, Styria, Austria. larschaefr@web.de.
Abstract:
This study aimed to evaluate whether continuous axillary temperature monitoring using a wearable patch enables earlier detection of postoperative infections compared to conventional intermittent infrared thermometry. 103 surgical patients were included in this prospective, single-center study and monitored over an 11-month period. Continuous axillary temperature monitoring using the SteadyTemp® patch was compared to routine infrared measurements performed as part of clinical routine. The primary outcome was fever detection rate (≥ 38.0 °C). Secondary outcomes included the correlation between fever detection and laboratory values as well as the frequency of clinical interventions. Out of 103 included patients, fever was detected in 33 cases. Continuous monitoring identified fever in 31 of these 33 patients (93.9%), whereas infrared thermometry detected fever in only 12 cases (36.4%). In 16 cases where antibiotic therapy was initiated or adjusted due to newly detected fever, the patch detected fever in 15 patients, compared to only 7 detections by infrared thermometry. Surgical interventions due to suspected infections were performed in 5 patients, and fever was detected by the patch in all cases, while infrared thermometry detected fever in only 2 of these patients. Due to the frequent failure of infrared thermometry to detect fever, a scoring system was developed to assess the clinical relevance of fever detection. Continuous temperature monitoring with the SteadyTemp® patch demonstrated superior fever detection compared to infrared thermometry, leading to earlier identification of febrile events. This study suggests that continuous temperature monitoring may enhance infection surveillance in surgical patients, allowing for more timely clinical interventions.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...