A Pipeline for 3D Multimodality Image Integration and Computer-assisted Planning in Epilepsy Surgery
Published on: May 20, 2016
Jointly Learned 3D Non-Cartesian Sampling With Wave Encoding and Reconstruction for Neurovascular Phase Contrast MRI
Chenwei Tang1,2, Brock W Jolicoeur2,3, James Rice2,4
1Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.
Accelerated 3D phase contrast (PC) MRI using learned wave encoding and model-based reconstruction (MoDL) significantly improves image quality and provides accurate flow measurements. This novel approach enables faster scans with comparable or better results than traditional methods.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging
- Fluid Dynamics
Background:
- Accelerating Magnetic Resonance Imaging (MRI) scans is crucial for reducing motion artifacts and improving patient comfort.
- 3D Phase Contrast (PC) MRI is essential for quantifying blood flow, but traditional methods are time-consuming.
- Wave encoding offers a promising strategy for accelerating MRI acquisition.
Purpose of the Study:
- To develop and validate an accelerated 3D PC MRI technique using jointly learned wave encoding and model-based reconstruction (MoDL).
- To compare the performance of the novel accelerated technique against conventional methods in terms of speed and accuracy.
Main Methods:
- A simulation framework and pseudo-fully sampled neurovascular 4D flow data were used to learn optimal wave encoding parameters and a MoDL network.
- Prospective 3D PC MRI scans were acquired in a flow phantom and 12 healthy participants using the learned wave encoding and a standard MoDL reconstruction.
- Acquisitions included the accelerated wave-encoded scan (2.25 min), a time-matched non-wave-encoded scan, a time-matched 3D radial scan, and a reference 3D radial scan (5.65 min).
Main Results:
- Learned wave scans in a phantom showed accurate flow rates and high correlation with reference scans (R² = 0.99).
- In vivo, wave-encoded scans exhibited reduced aliasing and blurring, with improved small vessel conspicuity compared to non-wave-encoded and time-matched radial scans.
- Quantitative flow analysis demonstrated similar or improved accuracy (coefficient of variation and intraclass correlation) for the accelerated wave scans compared to the reference scan.
Conclusions:
- Jointly learned wave encoding and MoDL reconstruction represent a feasible method for accelerating 3D PC MRI.
- The developed technique significantly enhances image quality and ensures accurate hemodynamic measurements.
- This acceleration holds potential for routine clinical application in neurovascular flow assessment.