Mechanical characterization of the Robert Jones bandage under cyclic compression-bending: An in vitro experimental
Samuel Larcheveque1, Stéphanie Claeys1, Marc Balligand1
1Teaching and Clinical Department of Companion Animals, Faculty of Veterinary Medicine, University of Liège, Liège, Belgium, Avenue de Cureghem, 1, 4000 Liège, Belgium.
Objective:
To assess the rigidity of Robert-Jones Bandage (RJB) used to stabilize a standardized in vitro fracture model, under cyclic compression/bending.
Study Design:
In vitro cyclic testing study.
Specimens:
wooden rods divided into two segments of identical length by a Cardan Joint enclosed in a thick foam tube and supported by a RJB.
Procedures:
Part 1: incremental loading applied to 6 specimens in successive stages of 50 N from 5 to 150 to 5-300 N with 3000 cycles per stage. Part 2: 9 specimens were loaded at 150 N (5-150 N) during 60,000 cycles.
Statistics:
The Wilcoxon signed-rank test was used to compare data obtained at 10,000 and 60,000 cycles.
Results:
In part 1, 6/6 specimens withstood 150 N for 3000 cycles. Three specimens endured an additional 3000 cycles at 200 N, 2 withstood 250 N for another 3000 cycles, and 1 tolerated 300 N. In part 2, 9/9 specimens withstood 150 N for 60,000 cycles, with displacements ranging from 0.102 mm to 0.309 mm (mean 0.179 ± 0.069 mm) and a median rigidity of 897 ± 375 N/mm. No specimen reached the fixed displacement limit of 2 mm after 60,000 cycles. No statistically significant difference was found between the results at 10,000 and 60,000 cycles (p = 0.129).
Conclusion:
The Robert-Jones Bandage showed consistent rigidity under prolonged cyclic loading in an in vitro fracture model.
Clinical Relevance:
While this in-vitro model cannot replicate the influence of soft tissues, joint motion, or multidirectional physiological forces, the findings suggest that a properly applied Robert Jones bandage may provide short-term mechanical support to a long-bone fracture prior to definitive stabilization. However, clinical applicability should be interpreted with caution given the inherent limitations of the experimental setup.
Related Concept Videos
Members Made of Elastoplastic Material
As the bending moment...
Normal Strain under Axial Loading
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Plastic Behavior
Residual Stresses in Bending