3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
Published on: November 27, 2017
Influence of Image Acquisition and Biologic Variability on Measurement Reproducibility of Vastus Lateralis Muscle
Enya T A Nacafucasaco1, Deivid D G Da Silva1, João G A Bergamasco1
1MUSCULAB-Laboratory of Neuromuscular Adaptations to Resistance Training, Department of Physical Education, Federal University of São Carlos (UFSCar), São Carlos, Brazil.
The single image acquisition protocol (1X) demonstrated superior reproducibility for vastus lateralis muscle architecture measurements compared to the two image acquisition protocol (2X). This highlights the importance of protocol selection in ultrasound assessments for reliable muscle analysis.
Area of Science:
- Biomechanics
- Musculoskeletal Ultrasound
- Exercise Physiology
Background:
- Accurate measurement of muscle architecture is crucial for understanding muscle function and adaptation.
- Ultrasonography (US) is a common tool for assessing muscle architecture, but its reproducibility can be influenced by various factors.
- Standardized protocols are essential for reliable and valid US-based muscle assessments.
Purpose of the Study:
- To compare the reproducibility of two distinct ultrasonography (US) protocols for measuring vastus lateralis muscle architecture.
- To evaluate the influence of single versus multiple image acquisitions on measurement variability.
- To determine the impact of protocol choice on key muscle architecture parameters: cross-sectional area (CSA), muscle thickness (MT), fascicle length (FL), and pennation angle (PA).
Main Methods:
- Two reproducibility protocols were compared: a single image acquisition (1X) and two separate image acquisitions (2X), both with 72-hour intervals.
- Muscle architecture parameters (CSA, MT, FL, PA) of the vastus lateralis were measured using US in 104 experimental units from 54 subjects.
- Reproducibility was analyzed using typical error (TE), coefficient of variation (CV), minimum detectable change (MDC), and Bland-Altman plots.
Main Results:
- The 1X protocol consistently yielded lower TE and CV values for all measured muscle architecture parameters compared to the 2X protocol.
- Significant differences between protocols were observed for fascicle length (FL) and pennation angle (PA) (p < 0.001 and p < 0.026, respectively).
- The 2X protocol exhibited greater variance across all parameters, indicating lower measurement reliability.
Conclusions:
- The single image acquisition (1X) protocol offers superior reproducibility for vastus lateralis muscle architecture assessment via ultrasonography.
- Protocol selection significantly influences measurement variability, impacting the reliability of muscle architecture data.
- These findings underscore the need to consider evaluator and biological variability when interpreting US-based muscle assessments.