Early detection of viable microorganisms in "sterile" periprosthetic fluids in implant-based breast reconstruction: A

Monari Marta1, Vaccari Stefano2, Lupacchini Alessandro Marco3

  • 1Laboratory of Clinical Analysis, Humanitas Mater Domini, Castellanza, Varese, Italy.

Abstract

Insights

A new bioelectrochemical method using electrochemical impedance spectroscopy (EIS) rapidly detects bacterial infections in breast reconstruction patients. This technique offers a faster, more accurate alternative to traditional methods for early infection diagnosis.

Area of Science:

  • Biomedical Engineering
  • Infectious Disease Diagnostics
  • Surgical Oncology

Background:

  • Postmastectomy implant-based breast reconstruction is susceptible to early postoperative infections.
  • Current diagnostic methods are time-consuming and may produce false negatives, especially after antibiotic use.
  • Early and accurate detection of bacterial periprosthetic fluid infections is crucial for effective treatment.

Purpose of the Study:

  • To evaluate the feasibility and diagnostic accuracy of electrochemical impedance spectroscopy (EIS) for early bacterial detection in periprosthetic fluid.
  • To assess EIS as a rapid, low-cost diagnostic tool for implant-based breast reconstruction infections.
  • To compare EIS performance against conventional microbiology and molecular diagnostics.

Main Methods:

  • Prospective enrollment of 47 patients undergoing immediate implant-based breast reconstruction.
  • Collection of 107 periprosthetic fluid samples at postoperative days 7 and/or 14.
  • Analysis of samples using EIS, molecular diagnostics, and conventional microbiology, with EIS interpreted against calibrated cut-offs.

Main Results:

  • EIS demonstrated full concordance with culture and molecular results, correctly identifying all 8 infected cases (17.0% incidence).
  • The technique differentiated gram-positive (Z' 6000-20,000 Ohms) from gram-negative infections (Z' 21,000-35,000 Ohms).
  • EIS analytical time was approximately 9 minutes; gram-negative infections were linked to severe outcomes.

Conclusions:

  • Electrochemical impedance spectroscopy (EIS) provides a rapid, cost-effective, and user-friendly method for detecting bacteria in periprosthetic fluids.
  • EIS bypasses the need for bacterial growth, sample preparation, or nucleic acid extraction, unlike standard tests.
  • Further validation of EIS holds significant potential for enabling earlier and more personalized interventions in breast reconstruction patients.