Electrochemical Roughening of Thin-Film Platinum Macro and Microelectrodes
Published on: June 30, 2019
Influence of Electrode Polishing Protocols, Potentiostat Models, and LOD Calculation Methods on the Electroanalytical
Michał Świderski1,2, Jagoda Seroka2, Dariusz Guziejewski2
1Doctoral School of Exact and Natural Sciences, University of Lodz, 90-237 Lodz, Poland.
Abstract:
The aim of this research is to present the extent to which the basic elements used in electrochemical measurements affect the results of electroanalytical procedures. Measurements were carried out using the square wave voltammetric technique on a glassy carbon electrode, and the recorded analytical signal corresponded to a model redox system. One of the objectives of the study was to illustrate the impact of using potentiostats from different manufacturers, as well as the variations observed among different models within the same brand. These models exhibited notable differences in both cost and advanced electrochemical measurement capabilities. The mechanical cleaning method for the solid disk electrode surface was also taken into consideration. Three different polishing motion types were tested, together with the number of repetitions. It was revealed that polishing motion significantly influences the electroactive surface area of the working electrode, as well as the repeatability of the measurements. The research showed that the largest electroactive surface area and the best repeatability of parameters are achieved when polishing is performed by drawing an 8-type motion on the polishing pad. The obtained results confirmed that the equipment and polishing applied in research have a greater than previously assumed impact on the statistical parameters characterizing the analytical procedure, for example, the limit of detection (LOD) or the dynamic range of the calibration curve. Both analyzed parameters have a significant impact on the quality of the statistical parameters describing derived analytical procedures. Finally, it was shown that significantly different statistical parameters can be obtained from the same set of data using various approaches for LOD estimation, with discrepancies reaching up to two orders of magnitude.
More Related Videos
Related Concept Videos
Voltammetry: Factors Affecting Measurements
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Standard Electrode Potentials
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...