A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
Dispersion-compensated Rowland spectrometer: implications for uranium VB-RIXS
Martin Sundermann1, Manuel Harder2, Ayman H Said3
1Deutsches Elektronen-Synchrotron DESY, Notkestr. 85, 22607 Hamburg, Germany.
Abstract:
The total energy resolution (ΔEtot) of a valence-band resonant inelastic X-ray scattering (VB-RIXS) instrument serves as an important point of reference in an otherwise complex field. Since VB-RIXS is a flux-limited technique, a pragmatic approach to reducing ΔEtot is often required-the specifications of a spectrometer should be matched with a comparable incident bandwidth (ΔEi) and the source size contribution (focal point) should be negligible. Although it advocates for a good efficiency, this approach is in many places already limited by count-rates. Here we follow a recent trend emerging in soft X-ray VB-RIXS and look at the performance of our tender X-ray Rowland spectrometer (Gretarsson et al., 2020) when being exposed to a source with a large linear dispersion (higher flux). Detailed ray tracing work, performed at the U M5-edge (3551 eV), finds that the intrinsic resolution of the Rowland spectrometer (ΔEa) can be obtained if the linear dispersion of the source matches the spectrometer's, but opposite in sign-here ΔEi does not matter. This finding is supported by experimental data where ΔEtot = 48 meV (ΔEa = 44 meV) was recently achieved. Furthermore, we demonstrate that the dispersion rate can be tuned, ensuring the method's applicability to other atomic edges.
Related Concept Videos
UV–Vis Spectrometers
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
UV–Vis Spectroscopy: Woodward–Fieser Rules
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
IR Spectrometers