A Microcontroller Operated Device for the Generation of Liquid Extracts from Conventional Cigarette Smoke and Electronic Cigarette Aerosol
Published on: January 18, 2018
Evaluation of carbonyl collection methods in electronic cigarette aerosols
Bianca Martinez1, Yeongkwon Son1, Andrey Khlystov1
1Organic Analytical Laboratory, Division of Atmospheric Science, Desert Research Institute, Reno, NV, United States.
Introduction:
This study evaluated the performance of three carbonyl collection media that are frequently used in electronic cigarette (e-cig) studies, namely: 2,4-Dinitrophenyl hydrazine (DNPH) silica impregnated cartridges (C), a DNPH-coated glass fiber filter (DF), and an impinger with a DNPH solution (I).
Methods:
These collection methods were tested with a 'mod' and a 'pod' e-cigs that represent devices producing different amounts of e-cig aerosols. The tests were performed using two different puff topography regimes to investigate the effect of puff flow rate on carbonyl collection efficiency. Carbonyls emitted from the e-cigs were collected using I-I (two impingers in series), C (a single DNPH cartridge), and DF-C (a DNPH filter followed by a DNPH cartridge). Samples were analyzed using high-performance liquid chromatography coupled with photodiode array detector.
Results:
Measured carbonyl levels varied between the three methods. For both devices, the highest values were measured with C and DF-C. Results from the 'pod' device measured with I-I were comparable with the other methods, but lower than the other two methods for the 'mod' device. For example, 1 LPM 'mod' acetaldehyde emissions measured with C (0.09 ± 0.02 µg/puff) and DF-C (0.14 ± 0.09 µg/puff) were significantly higher (p = 0.06 and 0.04, respectively) than I-I method (0.05 ± 0.02 µg/puff).
Discussion:
The likely cause for the underestimation by the I-I method is the inefficient collection of particle phase bound carbonyls by the impingers and, possibly, wall losses in the inlet of the impinger. Based on our results, C provides the best approach for carbonyl collection in e-cig aerosols, while I-I that is used in the standard testing method can underestimate carbonyl emissions when measuring high concentration aerosols such as those produced by 'mod' devices.
More Related Videos
08:39Generation of Electronic Cigarette Aerosol by a Third-Generation Machine-Vaping Device: Application to Toxicological Studies
Published on: August 25, 2018
08:33Use of Capillary Aerosol Generator in Continuous Production of Controlled Aerosol for Non-Clinical Studies
Published on: April 12, 2022
Related Concept Videos
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Atomic Absorption Spectroscopy: Atomization Methods