Validation and Subgroup Analysis of the Accuracy of the Photoplethysmography-based Microlife Cuffless Upper-arm
Ching-Fu Wang1, Han-Lin Wang2, Pei-Hsin Kuo3,4
1School of Biomedical Engineering, College of Biomedical Engineering, Taipei Medical University, Taipei City, Taiwan.
Background:
This study (1) validates the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure (BP) monitor according to the AAMI/ESH/ISO 81060-2:2018/Amd 1:2020 standard; (2) investigates the device's performance across age groups and skin tone; and (3) evaluates the advantages of cuffless BP monitoring in terms of estimation variability, required time, and stability.
Methods:
A total of 120 participants aged ≥20 years were recruited from the general population, ensuring the representation of diverse age and skin tone groups. The device was calibrated against a clinically validated mercury sphygmomanometer. Sequential same-arm BP estimation were performed. Accuracy was analyzed according to the ISO criteria, and further subgroup analysis was conducted to compare results between participants aged ≥65 and <65 years, as well as between participants with lighter and darker skin tones (classified according to the Fitzpatrick system). Additionally, estimation stability and variability were evaluated using 3 averaged BP estimations.
Results:
The results demonstrated that the Microlife device met the AAMI/ESH/ISO accuracy criteria. Subgroup analysis revealed consistent accuracy across age and skin tone groups, with slight differences warranting further exploration. Notably, the cuffless design enabled faster estimations with reduced variation between the 3 averaged readings, showcasing its potential for home BP monitoring and frequent self-assessments.
Conclusions:
These findings support the clinical potential of photoplethysmography-based cuffless BP monitoring in diverse adult populations, particularly for facilitating rapid and stable BP estimations in elderly individuals and users with varying skin tones. Further large-scale studies are warranted to corroborate and build upon these observations.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...