Related Experiment Video
Updated: Jan 8, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Determining optimal spot delivery pattern using log file Derived dose discrepancy factor in pencil beam scanning
Kah Seng Lew1, Calvin Wei Yang Koh2, Kang Hao Lee2
1Division of Radiation Oncology, National Cancer Centre Singapore, Singapore; Division of Physics and Applied Physics, Nanyang Technological University, Singapore.
Introduction:
Spatial uncertainties in spot position during reference dosimetry measurements of pencil beam scanning system were found to cause fluctuations in the reference dosimetry curve. This study aims to develop a method to determine the optimal delivery pattern for reference dosimetry measurement using information obtained from log file and quantifying the expected dose discrepancy.
Methods:
A total of 7 different delivery pattern were delivered five times each and log files for each delivery were collected. We introduced a new metric known as dose discrepancy factor (DDF) to quantify the expected dose discrepancy in the measurement arising from spot positioning error. DDF is calculated using a simple pencil beam algorithm and actual spot positions from the log file. Type A and B uncertainties of DDF are quantified to draw statistically rigorous conclusion on the most optimal delivery pattern.
Results:
Relative dose difference of up to 2 % can be observed in some of the delivery pattern when reference dosimetry curve was plotted. Linear fit of DDF calculated from log file against measured dose ratio yields an R2 of 0.56 which suggest moderate correlation. Delivery patterns with repainting and X fast scan (50MU) have the lowest average uncertainty. Friedman test and post-hoc test shows that different delivery patterns were proven to be statistical different with the exception of Y fast scan (50MU) versus both X fast scan (50MU) and random scan (50MU). Y fast scan (50MU) with the highest average DDF value of 0.979 ± 0.005 was found to be the most optimal delivery pattern as it has the least discrepancy from an ideal delivery.
Conclusion:
Delivery patterns used during reference dosimetry had a non-negligible impact on measurement. We had detailed a log file-based approach to determine the optimal delivery pattern such that dose measurement would be least affected by spot positioning error.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses