Related Experiment Video
Updated: Jan 8, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
X-Factor: Antithrombotic Coating Reduces Procedural Thromboembolic Events in Flow Redirection Endoluminal Device
M Harrison Snyder1, Alper Dincer1, Anthony J Yu1
1Cerebrovascular and Endovascular Division, Department of Neurosurgery, Tufts Medical Center, Boston, Massachusetts, USA.
Background:
The Flow Redirection Endoluminal Device (FRED) is a dual-layer flow diverting stent with a dense inner mesh and outer braided support providing superior deployment control. Covalently bonded Poly(2-methoxyethyl acrylate) X coating was introduced to decrease metal-induced thrombogenicity in the newer FRED X variant. The objective of the current study is to evaluate procedural thromboembolic events associated with both variants.
Methods:
A single-center retrospective analysis was undertaken comparing angiographically detected intraprocedural thromboembolic events, postprocedural magnetic resonance diffusion weighted imaging lesions (MR-DWI+), as well as a composite combined endpoint. Periprocedural (<24 hours) clinical events, demographics, aneurysm characteristics and platelet aggregation (using P2Y12 reactivity unit, PRU) were analyzed.
Results:
Treatment was performed with 19 FRED procedures in 18 patients with 19 aneurysms, and with 37 FRED X procedures in 35 patients harboring 40 aneurysms. Both groups achieved therapeutic preprocedural platelet inhibition by PRU testing. Intraprocedural thromboembolic events including intraluminal filling defects and side-branch slowing or loss occurred in 7/19 (36.8%) FRED and 0/37 FRED X procedures (P = 0.0002), with all being successfully recanalized using intra-arterial platelet inhibitors, without symptomatic ischemic or hemorrhagic events. The composite endpoint was reached in 11/19 (57.9%) FRED and 5/37 (13.5%) FRED X cases (P = 0.001), with multivariable analysis identifying device type (P = 0.015) among age, gender, and PRU.
Conclusions:
The surface modification of the FRED X device may be associated with improved thrombogenic risk profile in our cohort, with elimination of intraprocedural thrombus formation and reduction of composite thromboembolic events.
Related Concept Videos
Antiplatelet Drugs: Prostaglandin Synthesis, P2Y12 and Glycoprotein IIb/IIIa Inhibitors
Prostaglandin synthesis inhibitors, exemplified by the widely known aspirin, wield their power by irreversibly acetylating...
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Venous Thrombosis III: Interprofessional Care
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Peripheral Artery Disease III: Interprofessional Care