Novel Pacing Mode Conserves Battery in Dual-Chamber Leadless Pacemakers
James E Ip1, Nima Badie2, Mayer Rashtian3
1Weill Cornell Medicine/New York Presbyterian Hospital, New York, New York, USA.
Background:
A dual-chamber leadless pacemaker (LP) system has been developed, employing distinct atrial leadless pacemaker (ALP) and ventricular leadless pacemaker (VLP) devices that maintain atrioventricular synchrony via wireless implant-to-implant (i2i) communication with each beat. A new i2i communication-free programming option was introduced to extend battery longevity and minimize unnecessary ventricular pacing by providing rate-modulated atrial pacing with backup ventricular pacing when needed (AAI(R) + VVI). The clinical safety and longevity improvement have not been evaluated commercially.
Methods:
Patients indicated for dual-chamber pacing with new or existing dual-chamber LP implants were enrolled in this prospective, nonblinded, interventional study. Device diagnostics and battery longevity were interrogated before and after 6 months (6 M) of AAI(R) + VVI programming, with any programming-related complications or symptoms noted. Total ALP and VLP longevities were compared for DDD(R) at enrollment versus AAI(R) + VVI at 6 M.
Results:
Dual-chamber LP systems in 139 patients at six centers were evaluated (76 ± 11 years; 53% male; PR 210 ± 56 ms; 80% sinus node dysfunction, 16% AV block of any degree), including 16% new and 84% existing implants (enrolled 7.9 ± 7.5 months postimplant). The 84 patients completing the study (n = 55 did not return) demonstrated a complication-free rate of 100.0% and a symptom-free rate of 97.6% (n = 2; fatigue, dizziness, shortness of breath). In the 32 patients enrolled with existing implants already in DDD(R) mode who also completed the 6 M visit, AAI(R) + VVI extended the longevities of ALPs by 93% (5.8 ± 1.1 to 11.1 ± 2.7 years, p < 0.001) and VLPs by 61% (10.9 ± 2.1 to 16.9 ± 1.2 years, p < 0.001).
Conclusion:
Clinical use of AAI(R) + VVI programming in dual-chamber LPs over 6 M resulted in no complications, minimal symptoms, and significantly extended ALP and VLP battery longevities.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Heart Failure Drugs: Inotropic Agents
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Electrophysiology of Normal Cardiac Rhythm
Mechanism of Cardiac Arrhythmias
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...