Summary
Robotic-assisted total knee arthroplasty (TKA) showed superior radiographic alignment compared to conventional TKA. However, both methods yielded similar patient satisfaction and range of motion outcomes.
Area of Science:
- Orthopedic Surgery
- Robotics in Medicine
- Biomechanical Engineering
Background:
- Total knee arthroplasty (TKA) is a common procedure to address knee osteoarthritis.
- Advancements in surgical robotics offer potential improvements in precision and outcomes.
- The imageless Robotic Surgical Assistant (ROSA®) Knee System is one such technology for TKA.
Purpose of the Study:
- To compare the clinical and radiographic outcomes of conventional TKA versus robotic-assisted TKA.
- To evaluate the efficacy of the ROSA® Knee System in TKA procedures.
Main Methods:
- A prospective study comparing 23 conventional TKA patients with 27 robotic-assisted TKA patients.
- All surgeries performed by a single surgeon; radiographic measurements by two observers.
- Follow-up duration averaged 20.8 months for conventional and 17.6 months for robotic TKA.
Main Results:
- Robotic TKA showed significantly better radiographic alignment in coronal and sagittal planes (90.1% vs. 34.9%).
- Surgical time was longer for robotic TKA (78.3 min vs. 70.1 min).
- No significant differences in length of stay, patellar resurfacing, cementation, PCL resection, range of motion, or patient satisfaction were observed.
Conclusions:
- Robotic-assisted TKA using the ROSA® system demonstrates superior radiographic alignment compared to conventional TKA.
- Despite improved alignment, robotic TKA did not show significant advantages in range of motion or patient satisfaction.