Validation of the count-reduction method for planar bone scintigraphy: a phantom study focused on hot-lesion detection
Related Experiment Video
Updated: Jan 8, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Akinobu Kita1, Yoshihiro Nakamori2
1Department of Radiological Technology, Faculty of Health Sciences, Gifu University of Medical Science, 795-1 Nagamine, Ichihiraga, Gifu, Seki, 501-3892, Japan. akita@u-gifu-ms.ac.jp.
Abstract:
Poisson resampling can be used to synthesize low-count nuclear medicine images from full-count data. However, its fidelity relative to real short acquisitions remains uncertain. Using planar acquisitions of JSP and SIM2 bone phantoms filled with 99mTc, we compared the Poisson-generated and measured images at matched-count reductions (10–90%). The simulated total counts matched the measured counts at all reduction levels. The NMSE increased with a greater count reduction, and the values were similar between the groups at each level. In the bone phantom, the contrast ratio and CNR declined with fewer counts, whereas the %CV increased; the trends and values were similar between the simulated and measured images, even at a 90% reduction rate. These findings indicate that the Poisson-based count reduction can reproduce, without apparent systematic bias, the statistics and image-quality characteristics of true low-count images, supporting the safe and efficient optimization of low-dose and short-time protocols without additional patient scans.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015