Safety and Feasibility of a Fully Automated Insulin Delivery System: FCL@Home, a Multicenter Randomized Clinical
Marcela Moscoso-Vasquez1, Laya Ekhlaspour2, Sue A Brown1
1Center for Diabetes Technology, University of Virginia, Charlottesville, VA.
Objective:
Automated insulin delivery (AID) requiring manual premeal boluses are the standard of care for people with type 1 diabetes (T1D). We evaluated the safety and feasibility of the Automated Insulin Delivery as an Adaptive NETwork (AIDANET) system used without meal announcements in both supervised and home settings across a diverse population of people with T1D.
Research Design And Methods:
Adults (age >25 years), young adults (age 18-25 years), and adolescents (age 14-18 years) were enrolled at three sites in the U.S. to compare the AIDANET system used in fully closed-loop (FCL) with usual care (UC) in a randomized crossover study. Participants spent ∼5 days using FCL in a supervised environment followed by ∼7 days of use at home. The prespecified primary outcome was defined as the change in mean sensor glucose (MSG) between the 2nd week of UC and the week of the FCL at-home period.
Results:
Overall, 34 participants (12 adults, 10 young adults, 12 adolescents; mean age 25.4 ± 12.6 years; 62% female; mean HbA1c 8.0 ± 1.1%; 94% hybrid closed loop users during UC) completed the study. MSG improved from 178 to 164 mg/dL (UC vs. FCL P = 0.0058). The percentages of time in range (70-180 mg/dL), time in tight range (70-140 mg/dL), and time above range (>180 mg/dL and >250 mg/dL) also significantly improved, primarily overnight. The percentage of time <70 and <54 mg/dL were significantly noninferior in FCL compared with UC.
Conclusions:
FCL therapy with the AIDANET system can significantly improve average glycemic control while removing the need for mealtime bolusing for a broad population of people with T1D.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes Mellitus: Type 2 and Gestational
Hypoglycemia and Glucagon