Performance Verification of High Sensitivity Analyzer for TAT, PIC, TM, and t-PAIC
Yanhong Liu1, Bo Guo2, Guanghui Chen1
1Department of Laboratory Medicine, Henan Provincial People's Hospital, People's Hospital of Zhengzhou University, People's Hospital of Henan University, Zhengzhou, Henan, China.
This study validated the performance of four key hemostasis markers: thrombin-antithrombin complex (TAT), α2-plasmin inhibitor-plasmin complex (PIC), thrombomodulin (TM), and tissue plasminogen activator-plasminogen activator inhibitor complex (t-PAIC). Results show the HISCL-5000 analyzer is suitable for clinical diagnostics.
Area of Science:
- Clinical Chemistry
- Hematology
- Laboratory Medicine
Background:
- Thrombin-antithrombin complex (TAT), α2-plasmin inhibitor-plasmin complex (PIC), thrombomodulin (TM), and tissue plasminogen activator-plasminogen activator inhibitor complex (t-PAIC) are crucial in diagnosing thrombosis and hemorrhage.
- Current clinical practice lacks standardized performance verification methods for these four hemostasis markers.
Purpose of the Study:
- To evaluate the precision, trueness, carryover, linearity, and reference intervals of TAT, PIC, TM, and t-PAIC.
- To establish reliable performance standards for these indicators in clinical settings.
- To provide data and methodological references for future research.
Main Methods:
- Evaluation of precision, trueness, carryover, linearity, and reference intervals following CLSI guidelines (EP15-A2, EP06-A, C28-A).
- Analysis conducted on the HISCL-5000 analyzer.
- Verification intervals and manufacturer-assigned values were used for bias assessment.
Main Results:
- Within-laboratory coefficients of variation (CVs) were low for all four markers (e.g., TAT: 3.67-4.68%).
- Biases were within acceptable limits, and carryover rates were below 1%.
- Linearity was confirmed with correlation coefficients > 0.99, and manufacturer-provided reference intervals were appropriate.
Conclusions:
- The HISCL-5000 analyzer demonstrates acceptable performance for TAT, PIC, TM, and t-PAIC analysis.
- The system is suitable for clinical diagnostic applications, ensuring reliable hemostasis marker testing.