Related Experiment Video
Updated: Jan 8, 2026

Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Usefulness of metal artifact reduction on CT angiography after massive coil embolization in peripheral AVM
Jihwan Choi1, Sang Yub Lee2, Kwang Bo Park2
1Department of Digital Health, SAIHST, Sungkyunkwan University, 115 Irwon-Ro, Gangnam-gu, Seoul, the Republic of Korea; Department of Radiology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, the Republic of Korea.
Purpose:
To evaluate the image quality of three reconstruction methods-filtered back projection (FBP), adaptive statistical iterative reconstruction (AR50), and deep learning-based reconstruction (DL-M)-processed with the Smart Metal Artifact Reduction (SMAR) technique for post-embolization assessment of peripheral arteriovenous malformations (AVMs).
Materials And Methods:
In this prospective single-center study, 30 patients who underwent coil embolization for AVM were included. Post-embolization CT angiography was performed using dual-energy CT. Virtual monoenergetic images at 50 and 70 keV were reconstructed using FBP, AR50, and DL-M. All were processed with SMAR, and DL-M without SMAR served as the baseline. Artifact severity was objectively assessed using the standard deviation (SD) around the AVM, artifact index (AI), and contrast-to-noise ratio (CNR). Two readers (a resident and a staff radiologist) subjectively graded artifact severity, vessel visualization, and new artifacts using 4-point scales.
Results:
At both energy levels, average SD and AI were significantly lower in SMAR-processed images than in baseline DL-M (all p < 0.001). Subjective scores for artifact reduction and visualization of adjacent vessels were also significantly improved (p < 0.001). There were no significant differences among the three SMAR-processed methods. New artifacts appeared in three cases but had minimal effect on interpretability.
Conclusions:
SMAR-processed FBP, AR50, and DL-M reconstructions significantly reduced metal artifacts and improved visualization after AVM coil embolization, supporting their value for post-treatment evaluation and clinical decision-making.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...