Related Experiment Video
Updated: Jun 16, 2026

Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
Published on: August 14, 2018
Finite Element Investigation of Protective K-Wire Contribution to Hinge Fracture Prevention in Distal Femoral
Sohrab Rezaei1,2, Yamen Othmani3,4, Massamaesso Bilasse3,4
1ICube-Laboratoire des sciences de l'ingénieur, de l'informatique et de l'imagerie, UMR 7357, Université de Strasbourg, 4 Rue Blaise Pascal, 67000, Strasbourg, France. srezaeilafmejani@unistra.fr.
Purpose:
Distal femoral osteotomy (DFO) is an effective surgical procedure to correct valgus knee deformities. However, in lateral opening-wedge DFO (LOW-DFO), medial hinge fracture is a common complication that can lead to instability, correction loss, and delayed healing. This study aims to develop and validate a finite element (FE) model to investigate the mechanical effect of a protective Kirschner wire (K-wire) inserted in the medial hinge on the risk of hinge fracture.
Methods:
A 3D finite element model of the opening stage in LOW-DFO was developed based on a prior experimental study on 3D-printed femurs. Crack initiation and propagation leading to fracture were simulated using the extended finite element method (XFEM). Two scenarios were compared: one with a protective K-wire and one without. Characterization tests were performed to obtain the necessary mechanical properties.
Results:
The model was validated against data from previous mechanical experiments. The results showed that, while the protective K-wire had minimal effect on the location or onset of crack initiation, it increased the hinge stiffness and stabilized crack propagation. Compared to the configuration without the wire, the K-wire reduced the severity of unstable crack growth, thereby limiting abrupt failure.
Conclusions:
The use of a protective K-wire improves the mechanical stability of the medial hinge by moderating crack propagation without affecting the initiation threshold. This study highlights the importance of modeling crack propagation and not only initiation when evaluating fracture risk in DFO. The findings support the potential clinical benefit of K-wire insertion for preventing hinge-related complications.
More Related Videos
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025